The 15% kVp rule: how to trade kVp and mAs in radiography
This ventrodorsal radiograph of a cat was taken at 75 kVp and 5 mAs (data at the foot of the image) and the exposure was correct. The next patient is a cat of the same size that will not keep still, and you want to halve the exposure time to avoid blurring. Reference ventrodorsal radiograph:
zoom_inEnlarge
Which adjustment keeps a similar overall exposure?
chevron_right Show answer and explanation
A. Keep 75 kVp and double the exposure time to compensate
Doubling the time increases mAs and exposure, but worsens exactly what you wanted to avoid: motion blur.
B. Reduce kVp by 15% and halve the mAs
Lowering kVp by 15% is roughly equivalent to halving exposure; if you also halve the mAs, the image is clearly underexposed (noisy on digital systems).
C. Increase kVp by 15% (about 86 kVp) and halve the mAs (2.5 mAs)
Correct. Increasing kVp by 15% roughly doubles detector exposure, which compensates for halving the mAs. The image will have slightly less contrast (a longer grey scale), an acceptable trade-off for removing motion blur.
D. Double the kVp (150 kVp) and halve the mAs
kVp is not linearly related to exposure: doubling it grossly overexposes, increases scatter and destroys contrast. The practical rule is 15%, not double.
lightbulb Key takeaway
kVp controls energy (penetration and contrast) and mAs the number of photons (exposure and, on digital systems, noise). Rule of thumb: +15% kVp ≈ ×2 mAs. For the thorax use high kVp and low mAs with short times; for bone and abdomen, lower kVp for more contrast.
Test yourself with 10 random questions
Diagnostic imaging · 200 questions, a different test every time
Related questions